| Paul ion trap | 
|---|
| Paul ion trap Ion trapping device that depends on the application of radio frequency potentials between a ring electrode and two end-cap electrodes to confine the ion motion to a cyclic path described by an appropriate form of the Mathieu equation. The choice of these potentials determines the m/z value below which ions are not trapped. | 
| Related Term(s): | 
| Reference(s): O. D. Sparkman. Mass Spec Desk Reference, Global View Publishing, Pittsburgh (2006). R. E. March. J. Mass Spectrom. 32, 351 (1997). (http://dx.doi.org/10.1002/(SICI)1096-9888(199704)32:4<351::AID-JMS512>3.0.CO;2-Y ) J. F. J. Todd. Mass Spectrom. Rev. 10, 3 (1991). (http://dx.doi.org/10.1002/mas.1280100103 ) | 
| From Definitions of Terms Relating to Mass Spectrometry (IUPAC Recommendations 2013); DOI: 10.1351/PAC-REC-06-04-06 © IUPAC 2013. | 